Preliminary dosimetric evaluation of a designed head and neck phantom for intensity modulated radiation therapy (IMRT)
Keywords:
Intensity modulated radiation therapy (IMRT), verification, head and neck phantom, thermoluminescent.Abstract
The aims of this paper are to design, construct, and evaluate an anthropomorphic head and neck
phantom for dosimetric verification of nasopharyngeal cancer treatment plan using intensity modulated
radiation therapy (IMRT) technique. The phantom was designed as an assembly of thirty nine (39)
transversal section slabs fabricated from Perspex material each with delineated planning target
volumes (PTVs) and organ at risk (OARs) regions. Thermoluminescent dosimeter (TLD) was used after
multiple calibration cycles. The phantom was imaged, planned, and irradiated by IMRT plan. The
reproducibility of phantom measurements was checked by three identical IMRT irradiations. Four (4)
nasopharyngeal patients’ IMRT treatment plans were transferred to the phantom for dose verification.
Phantom’s measured doses were reproducible with less than 3.5% standard deviation. For the
verification of IMRT patient’s plans, the mean of percent dose differences between measured and
calculated doses was found 6.2% (SD: 4.7) at OAR and 5.96% (SD: 2.5%) at PTV. The percentage dose
deviation met the accuracy criteria of 7% at low dose regions. The standard deviation of TLD/TPS was
2.4% at PTV and 6.8% at OAR. This good agreement proves the feasibility of applying this phantom
in IMRT dose verification.